Aggregation of cell surface receptors is believed to be linked to their biological function. In this paper, we disclose the explicit formulation for macromolecular aggregation distribution by means of photobleaching Image Correlation Spectroscopy (pbICS).
Citation: Lajevardipour Alireza, W. M. Chon James, H. A. Clayton Andrew. Determining complex aggregate distributions of macromolecules using photobleaching image correlation microscopy[J]. AIMS Biophysics, 2015, 2(1): 1-7. doi: 10.3934/biophy.2015.1.1
Related Papers:
[1]
Albert W. Pilkington IV, Justin Legleiter .
Challenges in understanding the structure/activity relationship of Aβ oligomers. AIMS Biophysics, 2019, 6(1): 1-22.
doi: 10.3934/biophy.2019.1.1
[2]
O.S. Sorzano Carlos, Vargas Javier, Otón Joaquín, Abrishami Vahid, M. de la Rosa-Trevín José, del Riego Sandra, Fernández-Alderete Alejandro, Martínez-Rey Carlos, Marabini Roberto, M. Carazo José .
Fast and accurate conversion of atomic models into electron density maps. AIMS Biophysics, 2015, 2(1): 8-20.
doi: 10.3934/biophy.2015.1.8
[3]
Disha Mohan Bangalore, Ingrid Tessmer .
Unique insight into protein-DNA interactions from single molecule atomic force microscopy. AIMS Biophysics, 2018, 5(3): 194-216.
doi: 10.3934/biophy.2018.3.194
[4]
Oleksii V. Khorolskyi, Nikolay P. Malomuzh .
Macromolecular sizes of serum albumins in its aqueous solutions. AIMS Biophysics, 2020, 7(4): 219-235.
doi: 10.3934/biophy.2020017
[5]
Klemen Bohinc, Leo Lue .
On the electrostatics of DNA in chromatin. AIMS Biophysics, 2016, 3(1): 75-87.
doi: 10.3934/biophy.2016.1.75
[6]
David Kirchenbuechler, Yael Mutsafi, Ben Horowitz, Smadar Levin-Zaidman, Deborah Fass, Sharon G. Wolf, Michael Elbaum .
Cryo-STEM Tomography of Intact Vitrified Fibroblasts. AIMS Biophysics, 2015, 2(3): 259-273.
doi: 10.3934/biophy.2015.3.259
[7]
Ivana Sirangelo, Margherita Borriello, Gaetano Irace, Clara Iannuzzi .
Intrinsic blue-green fluorescence in amyloyd fibrils. AIMS Biophysics, 2018, 5(2): 155-165.
doi: 10.3934/biophy.2018.2.155
[8]
Marcelina Cardoso Dos Santos, Cyrille Vézy, Hamid Morjani, Rodolphe Jaffol .
Single cell adhesion strength assessed with variable-angle total internal reflection fluorescence microscopy. AIMS Biophysics, 2017, 4(3): 438-450.
doi: 10.3934/biophy.2017.3.438
[9]
Zachary Mekus, Jessica Cooley, Aaron George, Victoria Sabo, Morgan Strzegowski, Michelle Starz-Gaiano, Bradford E. Peercy .
Effects of cell packing on chemoattractant distribution within a tissue. AIMS Biophysics, 2018, 5(1): 1-21.
doi: 10.3934/biophy.2018.1.1
[10]
Mohamed A. Elblbesy .
Impact of erythrocytes properties on blood surface tension. AIMS Biophysics, 2019, 6(2): 39-46.
doi: 10.3934/biophy.2019.2.39
Abstract
Aggregation of cell surface receptors is believed to be linked to their biological function. In this paper, we disclose the explicit formulation for macromolecular aggregation distribution by means of photobleaching Image Correlation Spectroscopy (pbICS).
1. Introduction
The oligomeric state of cell surface receptors is believed to be linked to their biological functioning. Ligand-induced receptor oligomerisation appears to be a common mode of activation of tyrosine kinase receptors, such as the epidermal growth factor receptor. Large receptor clusters with as many as 20 receptors per cluster are implicated in the sensitive responses of bacteria to chemical gradients. Determining the precise nature of oligomeric states of cell surface receptors is a challenging biophysical problem. Although methods for water-soluble proteins are well established, methods for determining the complex oligomeric states of membrane proteins in cells are still on the rise.
Current methods of determining the brightness or average oligomeric state of macromolecules include fluorescence correlation spectroscopy [1], image correlation spectroscopy [2], photon counting histogram [3], spatial intensity distribution analysis [4], number and brightness [5], Forster resonance energy transfer [6] and fluorescence anisotropy [7] . These methods measure and analyse the fluctuations of fluorescence or the statistical properties of the fluorescence signals to yield estimates of average brightness or oligomeric state.
All of these methods require a brightness standard to relate the measurement to a reference oligomeric state or brightness standard. Moreover as is often the case for biophysical methods an average brightness or oligomeric state is extracted which precludes an understanding of complex aggregation or oligomeric distributions.
Single molecule step-wise photobleaching is one method that can extract oligomeric state information without a reference standard. The approach is based on the statistical bleaching properties of fluorophores. When a monomer bleaches there is a change in the intensity of emission from the monomer level to a background level. If there are two fluorophores in the region of interest, such as a dimeric protein, then after one bleach event the intensity will drop to the level of the monomer signal and then after a successive bleaching event the intensity will drop again to the level of the background signal. Identification of oligomers is possible without the requirement of a brightness standard and distributions from single oligomers are obtained which can be analysed to gain further information.
We have recently developed an analogous “ensemble” approach based on analysis of fluorescence images of labelled molecules obtained with standard (confocal) microscopy. The approach is called photobleaching image correlation microscopy and based upon the principle that the survival probability of an aggregate of molecules during photobleaching is dependent on the number of molecules in the aggregate. Consequently, a plot of the cluster density as a function of intensity remaining (not photobleached) after photobleaching provides a two-parameter estimate of aggregation or brightness. In our earlier paper we presented formulae for extracting the brightness or oligomeric state for a simple homogenous oligomer model.
In this report, we provide a more detailed derivation and present a list of explicit analytical formulae for complex distributions of molecules which are more likely to be encountered in real systems. We trust these formulae will find use for researchers interested in determining the oligomeric state distributions of molecules in complex environments such as biological cells.
2. Materials and Method
As photobleaching image correlation spectroscopy is a relatively new technique we will recall the salient features of the method for pedagogical purposes.
We begin by considering that our molecules of interest are labelled with an appropriate fluorophore (examples include GFP, or a dye such as fluorescein) and that the labelled molecules are presented in an appropriate form for imaging i.e. on a surface of an intact cell. For pbICS fixed cells are best because the method is suited for situations when molecules are not moving on the time-scale of the bleaching process.
By means of fluorescence microscopy one collects an image of the labelled molecules from the cells. A convenient form of microscopy is confocal microscopy because it allows a defined depth of focus and the zoom feature allows the image to be over sampled. This image is stored in memory. The process of image recording and saving is repeated until most of the original fluorescence is depleted.
The analysis of the images is done using a technique called image correlation spectroscopy.
2.1 Spatial Autocorrelation of Images-Image Correlation Spectroscopy
In an ergodic ensemble, the ensemble average of an stochastic process is equivalent to the spatial or temporal average. It empowers us to treat spatial and temporal averages correspondingly as ⟨I(t)⟩=⟨I(x)⟩. To apply this hypothesis in image analysis, the image should be homogeneous [8].
As Petersen et al.[2] showed, we can write 2D spatial autocorrelation for an image as:
G(η,σ)=⟨I(x,y)I(x+η,y+σ)⟩
(1a)
G(η,σ)=1MNM∑i=1N∑j=1I(i,j)I(i+η,j+σ)
(1b)
where M and N are the number of discrete points (pixels), η and σ represent spatial-lags for x and y, respectively. It is pragmatically important to express the autocorrelation of fluctuation function, δI(x,y)=I(x,y)−⟨I(x,y)⟩.
g(η,σ)=⟨δI(x,y)δI(x+η,y+σ)⟩⟨I(x,y)⟩2
(2a)
g(η,σ)=G(η,σ)⟨I(x,y)⟩2−1
(2a)
where ⟨I(x,y)⟩2
is normalising factor.
Practically, nobody use the above-mentioned equation to calculate autocorrelation of an image. Instead, one can exploit fast Fourier transformation along with the Weiner-Khinchin theorem to obtain the autocorrelation of the image more efficiently.
g(0,0)=FT−1{|FT{I(x,y)}|2}⟨I(x,y)⟩2−1
(3)
where FT represents Fourier transformation and |FT{I(x,y)}|2 is the power spectrum of the image. In practice, the maximum gray-value (zero-lag) of autocorrelation image can be used to calculate g(0,0) as
g(0,0)=The maximum gray-value of the autocorrelation imageMN×(The mean intensity of the image)2−1
(4)
2.2 General Theory of pbICS
Following statistical mechanics derivation for the fractional fluctuation of a random process in an ensemble [9] yields that the fractional fluctuation is proportional to the inverse number of particles in the observation volume:
⟨I−⟨I⟩⟩2⟨I⟩2=1⟨N⟩
(5)
where we have δI=I−⟨I⟩
g(0,0)=⟨δI⟩2⟨I⟩2=1⟨N⟩
(6)
then the autocorrelation function at zero-lag can provide us with the density of fluorescent molecules.
On the other hand, Elson and Magda [10] previously showed that there is direct connection between the detected intensity resulting from fluorescent fluctuation and the number of molecules of different species in illuminated volume.
I(t)=g∑jϵjQj∫L(r)Cj(r,t)d3r
(7)
where I(t) is detected intensity, L(r) displays laser intensity at point r, Cj(r,t)demonstrates the concentration of the jth fluorescent components at position r and time t, ϵj and Qj are molar extinction coefficient and fluorescent quantum yield of the j the component, respectively and g is an instrumental factor. Concentration fluctuation results in fluctuations in detected intensity:
δI(t)=g∑jϵjQj∫L(r)δCj(r,t)d3r
(8)
with some manipulation described by Elson and Magda [10], one can reach to:
where ⟨Cj⟩ is the mean concentration of the jth species. Therefore the detected intensity fluctuation is proportional to the concentration of fluorescent molecules.
Assuming that the molar extinction coefficient of a cluster linearly related to the molar extinction coefficient of its constituent monomers (i.e. ϵj=njϵ) and that quantum yield is unaffected by aggregation (Qj=Q) [11], then one can write [12]:
g(0,0)=∑jn2jcjV(∑jnjcj)2
(10)
where nj
is the number of monomers in the jth species and cj is the mean concentration (=⟨Cj⟩)
Since the photobleaching is a random process then the number of non-photobleached monomers in a specific species can fluctuate through sample, then we should use average number of monomers as:
g(0,0)=∑j⟨n2j⟩cjV(∑j⟨nj⟩cj)2
(11)
Finding average number of non-bleached monomers for jth species is connected to the historical problem in statistical mechanics [13]. The problem is similar to the classical problem of boxes containing white and black balls that solved by Newton for the first time. There are j boxes, each containing P white balls and Q black balls. We are interested to find the probability of finding i white balls by drawing one ball from each of the j boxes. We call the probability wj(i).
For drawing a single white ball from a box, the probability is obviously p=P/(P+Q), where it is q=1−p for a black ball. Generally, piqj−i is the probability of drawing i white balls from i specified boxes and j−i black balls from other j−i boxes. Including j!/(j−i)!i! ways to select i boxes from the j boxes, the answer will be
wj(i)=j!(j−i)!i!piqj−i
(12)
since p+q=1 and (a+b)n=n∑k=0n!(n−k)!k!akbn−k (binomial expansion) then j∑i=0wj(i)=(p+q)j=1. It means that wj(i) works as a probability density and we can use it to find mean value of variable as ⟨i⟩=j∑i=0iwj(i).
We consider the relation
(px+q)j=j∑i=0wj(i)xi
(13)
where x is an arbitrary variable. First and second derivatives with respect to x give
jp(px+q)j−1=j∑i=0iwj(i)xi−1
(14)
j(j−1)p2(px+q)j−2=j∑i=0i(i−1)wj(i)xi−2
(14)
If we let x=1 in these equations, the right hand sides give mean value of variables
jp=j∑i=0iwj(i)=⟨i⟩
(15a)
j(j−1)p2=j∑i=0i(i−1)wj(i)=⟨i2⟩−⟨i⟩
(15b)
In our case, after starting photobleaching process the number of non-bleached monomers will vary
⟨nj⟩=⟨i⟩=j∑i=0wj(i)i
(16a)
⟨n2j⟩=⟨i2⟩=j∑i=0wj(i)i2
(16b)
where nj for any j-mers with i non-bleached monomers equals ni with n=1 for a monomer. By applying equation (15), one can obtain
⟨nj⟩=jp,
(17a)
⟨n2j⟩=(jp)2+jp(1−p)
(17b)
then the equation (11) can be rewritten as
g(0,0)=∑j[(jp)2+jp(1−p)]cjV(∑jjpcj)2
(18)
By taking ⟨N⟩/V as cluster density, CD, and using equation (6) and (18) we will have
CD(p)=(∑jjcj)2p∑j[jcj+j(j−1)cjp]
(19)
That is the cluster density of a mixture of homogeneous j-mers as a function of fraction of remained fluorescence p. In other words, p is the probability of finding fluorescent components that can be calculated via dividing the mean intensity value of an image after bleach by the mean intensity value of the image before photobleaching. Ciccotosto and et al. [12] also introduced an explicit equation for a single j-mer that can be obtained by dropping summations in equation (19).
CDj(p)=jcjp1+(j−1)p,j=1,2,...
(20)
3. Results and Discussion
3.1 pbICS Explicit Aggregation Distribution
By using equation (19), we can obtain cluster densities of different mixtures of j-mers. For monomers with concentration of c1, dimers with concentration of c2 and so on, we have
where CD(p){1} is the cluster density of monomers with the mean concentration of c1, CD(p){1,2,4,6,8} demonstrates the cluster density of a mixture of monomers−dimers−tetramers−hexamers−octamers with mean concentrations of c1−c2−c4−c6−c8 respectively and p displays fraction of remained fluorescence. As expected, monomer density decays linearly with fraction of molecules remaining (not photobleached). However the pbICS curves from oligomers decay in a non-monotonic fashion.
For a specific ensemble, it is possible to fit Equation (21) for obtaining cjs with the general initial condition of
CDmax(p=1)=(∑jjcj)2∑jj2cj⋅
(22)
As a second method, it is noted that pbICS curves can also be fitted to the form of Equation (20) for finding average j with initial condition of cj=CDmax(p=1), where ⟨j⟩ is called the average brightness. Then the relation between ⟨j⟩ extracted from the latter method and cjs from the first approach can be formulated by
⟨j⟩=∑jj2cj∑jjcj=∑jjcjCDmax
(23)
Subsequently we can modify the Equation (20) to
CD(p)CDmax=⟨j⟩p1+(⟨j⟩−1)p⋅
(24)
3.2 Linear approximation
In some instances bleaching to completion may be difficult to achieve in practise and high doses of light exposure can be harmful to living cells. An alternative is to use gentle bleaching with a reduced power level such as obtained with replicate imaging under normal conditions. In this case one only partially bleaches the sample by perhaps 10\% or so. With a reduced range of $p$ values (now 1 to 0.9 instead of 1 to 0.1 that means 10\% photobleaching instead of 90\%) one cannot use the full non-linear Equation (19). Instead we propose to fit the data with a linear equation, which is essentially the tangent to the full function near p=1. For the homogenous model (or alternatively average brightness model), the linear form of the function is given by
CD(p)CDmax=(1−1⟨j⟩)+1⟨j⟩p,(p→1)
(25)
where ⟨j⟩ can then be obtained as ⟨j⟩=1/gradient or j=1/(1−intercept). Figure 1 reveals the full pbICS curves for j=1.5 and j=4.5 and the linear extrapolations near p=1.
Figure 1. The graph of Equation (24) and its linear approximation from Equation (25). The black line represents ⟨j⟩=1. The red curve and its linear approximation, red doted line, display ⟨j⟩=1.5 and the blue curve and dotted line depict ⟨j⟩=4.5.
We have elaborated upon the theoretical development of photobleaching image correlation microscopy. Formulae relating the pbICS curves to the oligomeric distribution were presented in closed form for the first time. In addition, a simple linear fitting method was introduced for small extents of photo-bleaching which should in turn find application in live cell applications.
5. Conflict of Interest
Authors report no conflicts of interest in this work.
References
[1]
Douglas Magde, Elliot Elson, W. Webb (1972) Thermodynamic Fluctuations in a Reacting System——Measurement by Fluorescence Correlation Spectroscopy.Phys Rev Let 19: 705-708.
[2]
N. O. Petersen, P. L. H?ddelius, P. W. Wiseman, et al. (1993) Quantitation of membrane receptor distributions by image correlation spectroscopy: concept and application.Biophys J 65: 1135-1146.
[3]
Yan Chen, Joachim D. Müller, Peter T.C. So, et al. (1999) The Photon Counting Histogram in Fluorescence Fluc- tuation Spectroscopy.Biophys J 77: 553-567.
[4]
Antoine G. Godin, Santiago Costantino, Louis-Etienne Lorenzo, et al. (2011) Revealing protein oligomerization and densities in situ using spatial intensity distribution analysis.Proc Natl Acad Sci U S A 108: 7010-7015.
[5]
Michelle A. Digman, Rooshin Dalal, Alan F. Horwitz, et al. (2008) Mapping the number of molecules and brightness in the laser scanning microscope.Biophy J 94: 2320-2332.
[6]
Tedman Torres, Marcia Levitus. (2007) Measuring conformational dynamics: a new FCS-FRET approach.J Phys Chem B 111: 7392-7400.
[7]
Joseph R. Lakowicz (2006) Principles of Fluorescence Spectroscopy.Springer US .
[8]
Claire Robertson and Steven C. George (2012) Theory and practical recommendations for autocorrelation-based image correlation spectroscopy.J Biomedical Optics 17: 80801-1.
[9]
Norman R. Davidson (1962) Statistical mechanics.McGraw-Hill .
[10]
Elliot L. Elson and Douglas Magde (1974) Fluorescence correlation spectroscopy. i. conceptual basis and theory.Biopolymers 13: 1-27.
[11]
N. O. Petersen (1986) Scanning uorescence correlation spectroscopy. i. theory and simulation of aggregation mea- surements.Biophys J 49: 809-815.
[12]
Giuseppe D. Ciccotosto, Noga Kozer, Timothy T. Y. Chow, et al. (2013) Aggregation distributions on cells determined by photobleaching image correlation spectroscopy.Biophys J 104: 1056-1064.
[13]
Henry Margenau, George Moseley MurphyThe mathematics of physics and chemistry.Young Press, 2nd edition1956 .
This article has been cited by:
1.
Andrew H A Clayton,
Fluorescence-based approaches for monitoring membrane receptor oligomerization,
2018,
43,
0250-5991,
463,
10.1007/s12038-018-9762-5
Laura C. Zanetti-Domingues, Dimitrios Korovesis, Sarah R. Needham, Christopher J. Tynan, Shiori Sagawa, Selene K. Roberts, Antonija Kuzmanic, Elena Ortiz-Zapater, Purvi Jain, Rob C. Roovers, Alireza Lajevardipour, Paul M. P. van Bergen en Henegouwen, George Santis, Andrew H. A. Clayton, David T. Clarke, Francesco L. Gervasio, Yibing Shan, David E. Shaw, Daniel J. Rolfe, Peter J. Parker, Marisa L. Martin-Fernandez,
The architecture of EGFR’s basal complexes reveals autoinhibition mechanisms in dimers and oligomers,
2018,
9,
2041-1723,
10.1038/s41467-018-06632-0
4.
Chiara Paviolo, James W. M. Chon, Andrew H. A. Clayton,
2018,
Chapter 4,
978-981-13-3064-3,
41,
10.1007/978-981-13-3065-0_4
5.
Alireza Lajevardipour, Andrew Clayton, Ammasi Periasamy, Peter T. So, Karsten König,
2019,
pbICS microscopy technique for determining oligomeric state,
9781510624061,
108,
10.1117/12.2508205
Lajevardipour Alireza, W. M. Chon James, H. A. Clayton Andrew. Determining complex aggregate distributions of macromolecules using photobleaching image correlation microscopy[J]. AIMS Biophysics, 2015, 2(1): 1-7. doi: 10.3934/biophy.2015.1.1
Lajevardipour Alireza, W. M. Chon James, H. A. Clayton Andrew. Determining complex aggregate distributions of macromolecules using photobleaching image correlation microscopy[J]. AIMS Biophysics, 2015, 2(1): 1-7. doi: 10.3934/biophy.2015.1.1
Figure 1. The graph of Equation (24) and its linear approximation from Equation (25). The black line represents ⟨j⟩=1. The red curve and its linear approximation, red doted line, display ⟨j⟩=1.5 and the blue curve and dotted line depict ⟨j⟩=4.5